| 摘要: |
| 目的 采用氨基酸分析仪测定依降钙素中氨基酸组成与肽含量,并与柱前衍生高效液相色谱法(HPLC法)测定结果比较。方法 色谱柱为2622SC-PH(4.6 mm×60 mm);A泵(pH系列溶液)流速为0.4 mL•min-1,B泵(茚三酮溶液)流速为0.35 mL•min-1;采用双波长检测(440 nm、570 nm)。结果 测得门冬氨酸等14种氨基酸的线性关系良好,重复性RSD为1.1%~2.5%,加样回收率为98.0%~103.6%,与HPLC法测定结果基本一致。结论 氨基酸分析仪法操作简单,自动化程度高,可用于定量分析依降钙素中氨基酸组成与肽含量。 |
| 关键词: 依降钙素 柱后衍生 阳离子交换色谱 氨基酸 肽 |
| DOI:10.13506/j.cnki.jpr.2025.02.011 |
| 基金项目:山东省重点研发计划项目(No.2022TZXD004);泰山产业领军人才工程项目(No.tscx202306076) |
|
出版日期:2025-02-25
|
|
| Determination of amino acid composition and peptide content in elcatonin by post-column derivatization cation-exchange chromatography |
| XUE Weili,LIANG Chaochao,SHI Feng,GONG Liping,ZHANG Dongmei,XIAN Ruiqing |
| (Shandong Institute for Food and Drug Control, NMPA Key Laboratory for research and Evaluation of Generic Drugs, Industrial Technology Foundation Public Service Platform, Jinan 250101, China) |
| Abstract: |
| Objective To determine the amino acid composition and peptide content in elcatonin using amino acid analyzer and compare the results with those obtained from the pre-column derivatization high-performance liquid chromatography (HPLC) method. Methods The chromatographic column was 2622SC-PH (4.6 mm×60 mm). The flow rate of pump A (pH series solution) was 0.4 mL•min-1, and the flow rate of pump B (ninhydrin solution) was 0.35 mL•min-1. The detection wavelengths were set at 440 nm (proline) and 570 nm (other amino acids). Results The results showed that the linear relationship of the 14 amino acids was good, the repeatability RSD was 1.1%-2.5%, and the recovery was 98.0%-103.6%. The results were generally consistent with those obtained from the HPLC method. Conclusion The amino acid analyzer method is simple and highly automated, and can be used to determine the amino acid composition and peptide content in elcatonin. |
| Key words: Elcatonin Post-column derivatization Cation-exchange chromatography Amino acids Peptides |