Cite this article:
XU Wendong,YU Peng,WANG Changcheng,MA Xiujuan,YUAN Jiangtao.Research on water distribution uniformity confirmation method of water bath sterilizer[J].Journal Of Pharmaceutical Research,2016,35(1):57-59[Click to copy]
【Print this page】      【Download PDF Full Text】   【View/Add Comment】  【Download reader】  【 Close 】
←Previous|Next→ Archive    Advanced Search
This article has been viewed: 3918times   Download 3880次 QR Code for this article
Scan with your phone!
水浴灭菌器布水均匀性确认方法研究
许文东1,2,于鹏2,王长城2,马秀娟2,原江涛2
Font:Larger+|Default|Smaller-
(1.山东大学药学院,山东 济南 250012;2.山东齐都药业有限公司,山东 淄博 255100)
摘要:
目的 建立一种确认水浴灭菌器布水均匀性的方法,对灭菌器布水均匀性进行评价。方法 通过对水浴灭菌器布水结构的分析,采用在布水口安装螺翼式水表对布水口的水量进行不同时间段的测量,对获得的数据进行显著性差异检验和相对标准偏差(RSD)分析,确认布水是否均匀。结果 通过对布水口单位时间的流量数据统计分析,显著性差异检验结果为15 min流量数据的P值为0.004(表示差异性极显著)、30 min流量数据的P值为0.963(表示差异性不显著)、60 min流量数据的P值为0.024(表示差异性显著); 相对标准偏差结果在2.3%至2.9%之间。结论 相对标准偏差对螺翼式水表获得的布水口水量监测数据进行统计评价为确认水浴灭菌器布水均匀性提供了简便和准确的方法。
关键词:  水浴灭菌器  布水均匀性  显著性差异检验  相对标准偏差
DOI:
基金项目:
Research on water distribution uniformity confirmation method of water bath sterilizer
XU Wendong1,2,YU Peng2,WANG Changcheng2,MA Xiujuan2,YUAN Jiangtao2
(1.School of Pharmaceutical,Shandong University,Jinan 250014,China;2.Shandong Qidu Pharmaceutical Co., Ltd.,Zibo 255400,China)
Abstract:
Objective To establish a water bath sterilizer confirmed the uniformity of water distribution method,and evaluate the uniformity of water distribution sterilizer.Methods Through the analysis of water bath sterilizer water distribution structure, the spiral wing type water meter installed in the nozzle of cloth on cloth outlet water in different time measurement,the obtained data was analyzed by significant difference test and RSD analysis,and to confirm that the water distribution was whether or not uniform.Results Through statistical analysis of the flow data of unit time,the significant difference test (test significance) results for P value of 15 min traffic data was 0.004(indicating that the difference is very significant), P value of 30 min flow data was 0.963 (indicating that the difference is not significant), P value of60min traffic data was 0.024(indicated significant difference); RSD results from 2.3% to 2.9%.Conclusion RSD of water monitoring data obtained from cloth outlet of spiral wing type water meter were statistically evaluated to confirm the water bath sterilizer water distribution uniformity for providing a simple and accurate method.
Key words:  Water bath sterilizer  Water distribution uniformity  Significant difference test  Relative standard deviation
Today's Visits:487    Total Visits:2383201
Website copyright:Editorial Board of Journal of Pharmaceutical Research     鲁ICP备17018036号-3     鲁公网安备 37010202001351号
Sponsored by:Shandong Institute for Food and Drug Control Shandong Pharmaceutical Association Address:No.2749 Xinluo Street,High-Tech Zone,Jinan,P.R.China Postcode:250101
Telephone Number:0531-81216586 Email :yaoxueyanjiu@163.com;yaoxueyanjiu@126.com
Technical Support:Beijing Qinyun Technology Co., Ltd.