• Home
  • About Journal
    Journal Introduction
    Editorial Board
    Youth Editorial Board
    Cooperating Institutions
    Editorial Department
  • Submission Guidelines
  • Policies & Statements
    Ethical Guidelines for Publication
    Academic Misconduct Statement
    Open Access (OA) Statement
    AIGC Use Statement
  • Journal Subscription
  • Download Center
  • 中文版
Cite this article:
XU Bin,MAO Ke,WU Xiaogang,CHEN Ning.Dissolution determination of Candesartan Cilexetil andAmlodipine Tablets[J].Journal Of Pharmaceutical Research,2017,36(10):585-588[Click to copy]
【Print this page】      【Download PDF Full Text】   【View/Add Comment】  【Download reader】  【 Close 】
←Previous|Next→ Archive    Advanced Search
This article has been viewed: 3735times   Download 2982次 QR Code for this article
Scan with your phone!
坎地沙坦酯氨氯地平片的溶出度测定
徐斌1,毛柯1,吴晓刚1,2,陈宁1
Font:Larger+|Default|Smaller-
(1.南京工业大学江苏省药物研究所有限公司,江苏 南京 210009;1.南京工业大学江苏省药物研究所有限公司,江苏 南京 210009;2.南京工业大学药学院,江苏 南京 210009)
摘要:
目的 建立坎地沙坦酯氨氯地平片中坎地沙坦酯及苯磺酸氨氯地平溶出度测定方法。方法按照中国药典2010年版(二部)附录Xc第二法(桨法),转速为每分钟50转,以1%吐温20水溶液为溶出介质,45 min时,对坎地沙坦酯取样;以0.01mol·L-1盐酸溶液为溶出介质,转速为每分钟50转,30min时对氨氯地平取样。采用C18色谱柱,流动相为甲醇-乙腈-20 mmol·L-1癸烷磺酸钠溶液[含0.04 mol·L-1磷酸二氢钾(磷酸调pH3.5)](600:100:300),检测波长238 nm,流速1.0 mL·min-1。结果坎地沙坦酯在0.562~8.992 ug·mL-1(r=0.9998)浓度范围内呈良好的线性关系,氨氯地平在0.292~4.672 µg·mL-1(r=0.9995)浓度范围内呈良好的线性关系;坎地沙坦酯、苯磺酸氨氯地平的平均回收率为101.0%±0.8%和 100.5%±0.9%。溶出度测定结果限度为坎地沙坦酯45min大于标示量的75%,苯磺酸氨氯地平30 min大于标示量的80%。结论所建立的方法专属性好,操作简便,可用于坎地沙坦酯氨氯地平片溶出度的检查。
关键词:  坎地沙坦酯  氨氯地平  高效液相色谱法  溶出度
DOI:10.13506/j.cnki.jpr.2017.10.008
基金项目:
Dissolution determination of Candesartan Cilexetil andAmlodipine Tablets
XU Bin1,MAO Ke1,WU Xiaogang1,2,CHEN Ning1
(1.Jiangsu Province Institute of Materia Medica,Nanjing University of Technology, Nanjing 210009,China;2.School of Pharmaceutical Sciences,Nanjing University of Technology,Nanjing 210009,China)
Abstract:
Objective To establish a dissolution method of candesartan cilexetil andamlodipine in Candesartan Cilexetil and Amlodipine Tablets.Methods According to the second dissolution method(paddle method)stated in the appendix of Chinese Pharmacopeia(2010 edition),dissolution medium of candesartancilexetil was purified water containing 1% polysorbate 20 and the rotating speed was 50 rpm,after 45 min, the dissolution solution was taken; Dissolution medium of amlodipine was 0.01mol·L-1 hydrochloric acid solution and the rotating speed was 50 rpm,after 30 min,the dissolution solution was taken.C18 column was used, mobile phase was that methanol- acetonitrile-20 mmol·L-1 decane sulfonate solution [containing 0.04 mol·L-1 KH2PO4 (phosphate tune pH 3.5)] (600:100:300), the detection wavelength was 238 nm and the flow rate was 1.0 mL·min-1.Results The linear range of candesartan cilexetil was 0.562 ~ 8.992 ug·mL-1(r=0.9998) and amlodipine was 0.292 ~4.672 ug·mL-1 (r=0.9995),the average recovery of candesartan cilexetil and amlodipine besylate were 101.0%±0.8% and 100.5%±0.9%,respectively.The dissolution of candesartan cilexetil was above 75% and amlodipine was above 80%.Conclusion The method was convenient and sensitive in the dissolution determination of Candesartan Cilexetil and Amlodipine Tablets.
Key words:  Candesartancilexetil  Amlodipinet  HPLC  Dissolution
Today's Visits:509    Total Visits:2383239
Website copyright:Editorial Board of Journal of Pharmaceutical Research     鲁ICP备17018036号-3     鲁公网安备 37010202001351号
Sponsored by:Shandong Institute for Food and Drug Control Shandong Pharmaceutical Association Address:No.2749 Xinluo Street,High-Tech Zone,Jinan,P.R.China Postcode:250101
Telephone Number:0531-81216586 Email :yaoxueyanjiu@163.com;yaoxueyanjiu@126.com
Technical Support:Beijing Qinyun Technology Co., Ltd.